Eddy-Current Sensors

ECL202 & ECL202e Eddy-Current Sensor: Probes and Ranges

Description down | Standard Ranges down | Fine Ranges down

Probe model numbers are specified by their diameter and body style: i.e. a U8C probe has an 8mm diameter and a "C" body style.

Probe diameter determines the range.
Larger Diameter = Larger Range

See the probe page for mechanical details, important mounting clearances, and temperature limits.

Probes can be ordered as vacuum compatible.

Ranges and Resolutions

Standard Range and Fine Range options are available.

A specific range and resolution will require a particular probe size/model. Range/Resolution may also be affected by the target material. "Standard" and "Fine" ranges are available for probes U3-U12. The Fine range has a smaller range but better resolution and are not available for the ECL202e.

Custom ranges are always available.

Listed specifications are for standard target materials.

Because of high resolutions, exports of the ECL202 may require an export license. The ECL202e has different resolution specifications and can ship without export license delays.

Specifications assume a target area at least 3 times the diameter of the probe.

cal

Standard Ranges, Resolutions, Thermal Drift

Range

mm
inch

Near Gap

mm
inch

Probe

Target
Type

ECL202
Resolution (RMS)*

µm
µinch

ECL202e
Resolution (RMS)*

µm
µinch

Thermal Drift**

±%F.S./°C

100 Hz 1 kHz 10 kHz 15 kHz 100 Hz 1 kHz 10 kHz 15 kHz Probe Driver
0.50
0.020
0.05
002

U3

nonferrous 0.025
1.0
0.030
1.2
0.060
2.4
0.065
2.6
0.300
12
0.300
12
0.300
12
0.300
12
0.04 0.04
ferrous 0.030
1.2
0.040
1.6
0.100
4.0
0.110
4.4
0.300
12
0.300
12
0.300
12
0.300
12
0.08 0.08
1.25
0.050
0.25
010

U5

nonferrous 0.045
1.8
0.065
2.5
0.140
5.5
0.150
6.0
0.300
12
0.300
12
0.300
12
0.300
12
0.04 0.10
ferrous 0.080
3.2
0.120
4.8
0.240
9.5
0.260
10.5
0.300
12
0.300
12
0.300
12
0.300
12
0.10 0.10
20
0.080
0.35
015

U8

nonferrous 0.040
1.6
0.060
2.4
0.135
5.5
0.145
6.0
0.300
12
0.300
12
0.400
16
0.400
16
0.02 0.04
ferrous 0.070
3.0
0.080
4.0
0.180
7.0
0.200
8.0
0.300
12
0.300
12
0.400
16
0.400
16
0.04 0.04
3.50
0.140
0.60
025

U12

nonferrous 0.060
2.5
0.090
3.5
0.210
9.0
0.240
10
0.300
12
0.300
12
0.400
16
0.400
16
0.02 0.01
ferrous 0.100
4.0
0.170
7.0
0.250
10
0.300
12
0.300
12
0.300
12
0.400
16
0.400
16
0.03 0.01
50
0.200
0.75
030

U18

nonferrous 0.080
3.0
0.130
5.0
0.300
12
0.340
14
0.400
16
0.400
16
0.500
20
0.500
20
0.01 0.01
ferrous 0.130
5.0
0.200
8.0
0.390
16
0.450
18
0.400
16
0.400
16
0.500
20
0.500
20
0.01 0.01
80
0.320
1.25
050

U25

nonferrous 0.180
7.0
0.250
10
0.500
20
0.600
24
0.400
16
0.400
16
0.550
22
0.600
24
0.01 0.01
ferrous 0.180
7.0
0.250
10
0.500
20
0.600
24
0.400
16
0.400
16
0.600
24
0.650
26
0.01 0.01
12.5
0.500
1.50
060

U38

nonferrous 0.200
8.0
0.350
14
0.700
28
0.800
32
0.500
20
0.500
20
0.700
28
0.750
30
0.01 0.01
ferrous 0.200
8.0
0.350
14
0.700
28
0.800
32
0.500
20
0.500
20
0.750
30
0.800
31
0.02 0.01
15
0.600
20
080

U50

nonferrous 0.300
12
0.400
16
0.800
32
0.900
36
0.600
24
0.600
24
0.750
30
0.800
31
0.01 0.01
ferrous 0.300
12
0.450
18
0.900
36
1.00
40
0.600
24
0.600
24
0.850
33
0.900
35
0.01 0.01

* Peak-to-Peak values are approximately 10X RMS values
In high EMI environments, output noise may rise to 0.030VRMS (0.3% resolution)

** Temperature coefficient specified at: Probe: 15°C-65°C; Driver: 15°C-65°C

 

Fine Ranges (Smaller Ranges, Higher Resolutions)

Range

mm
inch

Near Gap

mm
inch

Probe

Target
Type

ECL202
Resolution (RMS)*

µm
µinch

Thermal Drift**

±%F.S./°C

100 Hz 1 kHz 10 kHz 15 kHz Probe Driver
0.25
0.010
0.05
0.002

U3

nonferrous 0.010
0.4
0.015
0.6
0.025
1.0
0.030
1.2
0.04 0.05
ferrous 0.015
0.6
0.025
1.0
0.040
1.6
0.050
2.0
0.06 0.10
0.625
0.025
0.25
0.010

U5

nonferrous 0.030
1.2
0.035
1.4
0.070
2.8
0.075
3.0
0.04 0.04
ferrous 0.040
1.6
0.050
2.0
0.075
3.0
0.080
3.2
0.08 0.04
1.0
0.040
0.35
0.015

U8

nonferrous 0.020
0.8
0.030
1.2
0.050
2.0
0.060
2.4
0.02 0.04
ferrous 0.050
2.0
0.060
2.4
0.100
4.0
0.110
4.3
0.04 0.04
1.6
0.065
0.60
0.025

U12

nonferrous 0.040
1.6
0.050
2.0
0.100
4.0
0.110
4.4
0.01 0.01
ferrous 0.050
2.0
0.070
2.8
0.120
4.8
0.130
5.1
0.02 0.02
* Peak-to-Peak values are approximately 10X RMS values
In high EMI environments, output noise may rise to 0.030VRMS (0.3% resolution)

 

 