ECL101/ECL110 Wide Bandwidth
ECA101 Analog Proximity ECD140 cRIO, High-Performance ECA110 OEM PCB BasedCUSTOM Electronics Probe Details Thread Detection Probes Accessories Discontinued Products FAQECL101 Documents
Application Notes
Measure Micro-Displacement Position Measurement Thread Detection More Application NotesLinear output at up to 80 kHz bandwidth. Can be calibrated in the field.
Features
| ECL110 MultiChannel
| Specifications
| Probes & Ranges
| Mechanical
Life-Cycle Status: Active
Thread Sensing in Nonferrous Materials
ECL110
The ECL110 is a multi-channel, OEM version of the ECL101. The ECL110 has no enclosure, is powered by a single power connection, and can have 1-8 channels internally synchronized with no need for external sync cables.
Typical specifications for standard targets and probes. Actual values depend on probe, target, and calibration. Specifications assume a target area at least three times the diameter of the probe.
Resolution: |
See Range Table for resolution specifications |
|
Linearity |
±0.50% of Range | |
Bandwidth (Factory Set) |
1 kHz, 10 kHz, 80 kHz | |
Temperature Drift of Probe at mid-range |
±0.02% of Range/°F ±0.04% of Range/°C |
|
Output |
0-10 VDC | |
Input Power |
12-24 VDC 130 mA at 15 VDC |
|
Operating Temperature |
40°F-120°F 4°C-50°C |
|
Probe Operating TemperaturesStandard Probes come with PUR (Polyurethane) jacketed cables. High Temperature Probes have FEP jackets. Other temperature ranges may be possible. Contact us for more information. Standard Probes: IP67 High Temperature Probes: IP63 |
U3 Probe |
Standard Probe (PUR Cable): High Temperature Version Not Available |
| U5-18 Probes | Standard Probes: High Temperature Probes: |
|
| U25-50 Probes | Standard Probes: High Temperature Probes: |
|
Larger Probe Diameter = Larger Range Probes can be ordered as vacuum compatible; custom probes are available. |
|
Listed specifications are for standard target materials: Nonferrous Al 6160, Ferrous 4140 Steel.
Range mm |
Near Gap mm |
Probe |
Target |
Resolution (RMS*) µm |
||
| 1 kHz | 10 kHz | 80 kHz | ||||
| 0.50 0.020 |
0.05 0.002 |
U3 |
Nonferrous | 0.03 1.2 |
0.06 2.4 |
0.20 7.9 |
| Ferrous | 0.04 1.6 |
0.08 3.1 |
0.30 12 |
|||
| 1.25 0.050 |
0.25 0.010 |
U5 | Nonferrous | 0.06 2.4 |
0.10 4.0 |
0.25 9.8 |
| Ferrous | 0.09 3.5 |
0.15 5.9 |
2.0 79 |
|||
| 2.00 0.080 |
0.35 0.015 |
U8 | Nonferrous | 0.10 4.0 |
0.16 6.3 |
1.2 47 |
| Ferrous | 0.13 5.1 |
0.21 8.3 |
2.0 79 |
|||
| 3.50 0.140 |
0.60 0.025 |
U12 | Nonferrous | 0.20 7.9 |
0.28 11 |
0.70 28 |
| Ferrous | 0.26 10 |
0.35 14 |
1.2 47 |
|||
| 5.00 0.200 |
0.75 0.030 |
U18 | Nonferrous | 0.24 9.4 |
0.48 19 |
3.2 130 |
| Ferrous | 0.32 13 |
0.64 25 |
4.5 180 |
|||
| 8.00 0.320 |
1.25 0.050 |
U25 |
Nonferrous | 0.35 14 |
0.70 28 |
5.3 210 |
| Ferrous | ||||||
| 12.5 0.500 |
1.50 0.060 |
U38 | Nonferrous | 0.55 22 |
1.1 43 |
8.3 330 |
| Ferrous | ||||||
| 15.0 0.600 |
2.00 0.080 |
U50 | Nonferrous | 0.66 26 |
1.3 51 |
10 390 |
| Ferrous | ||||||
| * Peak-to-Peak values are approximately 8-10 times RMS values In high EMI environments (10 V/m), output noise may rise to 0.050VRMS (0.5% resolution) |
||||||



Click for mechanical details

Click for mechanical details

Click for mechanical details

Click for mechanical details

Click for mechanical details

Click for mechanical details

Click for mechanical details

Click for mechanical details