Eddy-Current Sensors

ECL101/ECL110 Eddy-Current Sensor: Probes and Ranges

Probe model numbers are specified by their diameter and body style: i.e. a U8C probe has an 8mm diameter and a "C" body style.

Probe diameter determines the range.
Larger Diameter = Larger Range

See the probe page for mechanical details, important mounting clearances, and temperature limits.

Probes can be ordered as vacuum compatible.

Range and Resolution

Specifications assume a target at least 3 times the diameter of the probe.

A specific range and resolution will require a particular probe size/model. Range/Resolution may also be affected by the target material.

Custom ranges are always available.

Listed specifications are for standard target materials.

Range and Resolution Chartcal

Range

mm
inch

Near Gap

mm
inch

Probe

Target
Type

Resolution (RMS*)

µm
µinch

1 kHz 10 kHz 80 kHz
0.50
0.020
0.05
0.002
U3 Nonferrous 0.03
1.2
0.06
2.4
0.20
7.9
Ferrous 0.04
1.6
0.08
3.1
0.30
12
1.25
0.050
0.25
0.010
U5 Nonferrous 0.06
2.4
0.10
4.0
0.25
9.8
Ferrous 0.09
3.5
0.15
5.9
2.0
79
2.00
0.080
0.35
0.015
U8 Nonferrous 0.10
4.0
0.16
6.3
1.2
47
Ferrous 0.13
5.1
0.21
8.3
2.0
79
3.50
0.140
0.60
0.025
U12 Nonferrous 0.20
7.9
0.28
11
0.70
28
Ferrous 0.26
10
0.35
14
1.2
47
5.00
0.200
0.75
0.030
U18 Nonferrous 0.24
9.4
0.48
19
3.2
130
Ferrous 0.32
13
0.64
25
4.5
180
8.00
0.320
1.25
0.050
U25 Nonferrous 0.35
14
0.70
28
5.3
210
Ferrous
12.5
0.500
1.50
0.060
U38 Nonferrous 0.55
22
1.1
43
8.3
330
Ferrous
15.0
0.600
2.00
0.080
U50 Nonferrous 0.66
26
1.3
51
10
390
Ferrous
* Peak-to-Peak values are approximately 8-10 times RMS values