Eddy-Current Sensors
ECD140 & ECD140e Eddy-Current Sensor for CompactRIO:
Probes and Ranges
Probe model numbers are specified by their diameter and body style: i.e. a U8C probe has an 8mm diameter and a "C" body style.
Probe diameter determines the range.
Larger Diameter = Larger Range
See the probe page for mechanical details, important mounting clearances, and temperature limits.
Probes can be ordered as vacuum compatible.
Ranges and resolution
Specifications assume a target at least 3 times the diameter of the probe.
A specific range and resolution will require a particular probe size/model. Range/Resolution may also be affected by the target material.
Custom ranges are also available.
Listed specifications are for standard target materials.
Terms

Rangemm |
Near Gapmm |
Probe |
Target |
ECD140µm |
ECD140e
µm |
| 2.00 |
0.35 |
U8 | nonferrous | 0.145 |
0.4 |
| ferrous | 0.180 |
0.4 |
|||
| 3.50 |
0.60 |
U12 | nonferrous | 0.240 |
0.4 |
| ferrous | 0.390 | 0.390 | |||
| * RMS @ 15 kHz. Peak-to-Peak values are approximately 8-10 times higher In high EMI environments, output noise may rise to 0.030 VRMS (0.3% resolution) |
|||||