PDF Range Chart with more detail (66k)

Probes
Range, Resolution, Linearity*
Sensing
Area
Body Style Options Ultrafine (Elite Only) Fine Standard Extended (Elite Only)
Range
µm
mils
Resolution Lin.
%FS
Range
µm
mils
Resolution Lin.
%FS
Range
µm
mils
Resolution Lin.
%FS
Range
µm
mils
Resolution Lin.
%FS
BW
Hz
nm
µin
BW
Hz
nm
µin
BW
Hz
nm
µin
BW
Hz
nm
µin
F** C3 Probe C5 Probe C6 Probe - - - - 10
0.4
100 1.1
0.04
0.25 50
2.0
100 3.8
0.15
0.30 80
3.0
100 6.8
0.27
0.85
- - 6k 1.5
0.06
6k 5.0
0.20
6k 9.0
0.36
E** C3 Probe C5 Probe C6 Probe - - - - 25
1.0
100 3.0
0.12
0.15 100
4.0
100 7.5
0.30
0.15 - - - -
- - 6k 4.0
0.16
6k 10
0.40
- -
D C3 Probe C5 Probe C6 Probe 10
0.4
100 0.2
0.01
0.15 50
2.0
100 1.0
0.04
0.15 250
10.0
100 4.5
0.18
0.10 500
20.0
100 9.0
0.36
0.15
15k 0.4
0.02
15k 2.0
0.08
15k 9.0
0.35
15k 18
0.71
C C23 Probe C24 Probe - 10
0.4
100 0.2
0.01
0.20 50
2.0
100 1.2
0.05
0.15 250
10.0
100 5.0
0.20
0.10 500
20.0
100 9.0
0.36
0.20
15k 0.4
0.02
15k 2.4
0.10
15k 10
0.40
15k 18
0.71
B C23 Probe C24 Probe - - - - - 50
2.0
100 1.0
0.04
0.20 500
20.0
100 5.0
0.20
0.15 1250
50.0
100 20
0.79
0.20
- - 15k 2.0
0.08
15k 10
0.40
15k 40
1.6
A C1 Probe C2 Probe - - - - - 50
2.0
100 0.8
0.03
0.20 500
20.0
100 6.0
0.24
0.20 2000
80.0
100 20
0.79
0.45
- - 15k 1.6
0.06
15k 12
0.47
15k 40
1.6
H R2 Probe - - - - - - - - - - 2500
100.0
100 50
2.0
0.20 6000
250.0
100 90
3.5
0.25
- - - - 15k 100
4.0
15k 180
7.1

*Resolution and linearity (lin.) values are for Elite Series sensors. Compact Driver resolution and linearity values ar 2X higher.
For other bandwidths use 100Hz values with these multipliers: 1kHz-1.4, 10kHz-1.8, 15kHz, 2.0 (15kHz is standard/unfiltered)
Resolution values are RMS. Peak-to-peak values are typically ten times greater than the RMS values.
**10kHz maximum bandwidth for these sensing areas.